Т³гж 25 – LABEC AA-17BS User Manual
Page 24

Operation manual
SEA Series Steam Sterilizers
20
Fig 9-10
SELECT A CYCLE
CLEAN GENERATOR
CLEAN GENERATOR
Fig9-11
Please close door
Fig9-12
CLEAN REQ.
NOTICE
9.7 Cleaning the Steam Generator
1. Purpose:
To clean the steam generator and steam pipes, to prevent
furring blocking the pipe.
2. Cycle:
20 min cleaning + 10 min drying
3. Turn on the autoclave via the ON / OFF switch. The screen
will display the welcome state.
4. Cleaning Cycle Operation:
Press the DRY key 6 times until CLEAN GENERATOR is
displayed. Press OK.
5. Close the door and press START. The autoclave will now
start the cleaning cycle.
6. Cleaning Reminder:
Every 100 cycles the B&D light and LEAK light will go on to
remind you to clean the steam generator.
9.8 Repair and change of parts must be done by qualified service engineer from the
manufacturer or an authorized contractor.
9.9 If the machine needs to be maintained or the electric spare parts need to be changed please
cut off the power first and ask your dealer or manufacturer to change it.
Cleaning function will not affect the autoclave cycle,
only a remind message.
Note: