KEEP WARMING

UNIVERSAL B Cycle

Fig5-2

Please close door

5.3 Select cycle

Press MENU/C key to display cycle options. There are 7 cycles displayed
in three page layouts (Fig 5-2).
Press STEP key and QUICK key to select cycle.
Press OK key to confirm selection. You need to confirm cycle
choice by pressing OK otherwise "Invalidation" will be displayed.

CYCLES
Universal B cycle:
134

Three times pre-vacuum, 4 min sterilization time

18 min B cycle:
134

Three times pre-vacuum, 18 min sterilization time.

Suitable for instruments that need a long sterilization
time. It can achieve the best sterilization effect.

Unwrapped cycle:
134

one time pre-vacuum, 4 min sterilization time.

Rubber & Plastic cycle:
121

Three times pre-vacuum, 16 min sterilization time.

B & D Testing cycle:
BOWIE & DICK, Leak Test.

Vacuum test cycle:
It is used to test the chamber's sealing property during the
vacuum test cycle.

Cleaning steam generator:
Clean the steam generator and steam pipe.
Refer to Page 19 for specific operation.

UNIVERSIAL B CYCLE

18 MINUTES B CYCLE

UNWRAPPED CYCLE

SELECT A CYCLE:

RUBB & PLAS.CYCLE

LEAK TEST CYCLE

B&D TEST CYCLE

SELECT A CYCLE:

CLEAN GENERATDR

SELECT A CYCLE:

Fig5-2

Operation manual

SEA Series Steam Sterilizers

7