KERN & SOHN AEJ 100-4NM User Manual
Page 68

68
AEJ_N/AES_N-BA-e-1321
13.6.3 Carry out series of measurements for test volume V
1/2
Carry out the second series of measurements by following the same sequence
of operations as the first, see chap. 13.6.2
1.
Dispense the test fluid (such as distilled water) into the weighing vessel of
conduit “V
1/2
” to be measured.
2.
Wait for stability display, then press the PRINT button. Place 2 followed by
the current weight display will appear.
3.
Tare, by pressing the TARE-key.
Repeat this test cycle (steps 1 -3) as often as set in “Set n”
(See chap. 13.6.1).
After the last test cycle has been confirmed a display announcing the start of
the third series of measurements “P2 V
1
”, followed by Place 1, see chap.
13.6.4, will appear.