4 weighing units switch-over, 5 taring – KERN & SOHN AEJ 100-4NM User Manual
Page 36
![background image](/manuals/697268/36/background.png)
36
AEJ_N/AES_N-BA-e-1321
Confirm your selection by pressing the PRINT-key; weighing scale will return to
menu.
For returning to weighing mode with set weighing unit, see chap. 8.2.
9.4 Weighing units switch-over
A weighing unit may be changed by pressing the UNIT-key several times:
[g]
→ [mg] → [ct] → [oz] → [ozt] → [dwt] → [t] → [mom]→ [G] → [g]
Note:
In calibrated model selection is restricted to options [g]
→ [mg] → [ct].
9.5 Taring
The dead weight of any weighing container may be tared away by pressing a button, so that
the following weighing procedures show the net weight of the goods to be weighed.
Put vessel of goods to be weighed on weighing plate and close the wind
screen doors.
Wait until the stability display appears (
), then press TARE. The weight of
the container is now internally saved.
Weigh the goods to be weighed and close the wind screen doors.
Wait until the stability display appears (
).
Read net weight.
• When the balance is unloaded the saved taring value is displayed with
negative sign.
• To delete the stored tare value, remove load from weighing pan and
press TARE.
• The taring process can be repeated any number of times. The limit is
reached when the whole weighing range is exhausted.