3 menu settings “p4 print – KERN & SOHN AEJ 100-4NM User Manual
Page 73
AEJ_N/AES_N-BA-e-1321
73
14.3 Menu settings “P4 PRINT“
Menu call “P4 PRINT“, see chap. 8.1.
Menu overview
Function
Settings
Description of options
P4.1 bAud
2400 bit/s
Transfer speed
4800 bit/s *
9600 bit/s
19200 bit/s
P4.2 CntA
no*
Continuous output disabled in standard
weighing unit,
Output requires key operation
YES
Continuous output in standard weighing unit
P4.3 Cntb
no*
Continuous output disabled in current weighing
unit, output requires key operation
YES
Continuous output in current weighing unit
P4.4 rEPL
no*
Manual issue after pressing the PRINT key.
YES
Automatic issue of first stable weighing value
Sequence of operations:
1. Taring
2. Place weight, issue of first stable weighing
value
3. Renewed output only possible after weight
was removed. Prerequisite: Display +/- 50
display steps from zero.
4. Place next weight.
P4.5 PStb
Locked in models with
type approval certificate
no
Output even for unstable weighing value
YES*
Output for stable weighing value only
P4.6 Lo
0.01*
Input minimum weight for automatic output:
Weighing value issued automatically if current
weighing value exceeds entered minimum
value. The next weighing value will not be
issued unless the weighing value has
meanwhile dropped below the entered weighing
value.
Factory settings are marked by *.