7.1 Automatic adjustment using internal weight (AEJ only)
With the internal adjustment weight, the weighing accuracy can be checked and re-
adjusted at any time.
The automatic adjustment function is always enabled. You can start adjustment at
any time by pressing the F-key 3 times.
Temperature / time controlled adjustment is taking place:
Automatic adjustment will be indicated 5 minutes in advance by “°C” (change of
temperature
∆t 0.8°C) or the “►”symbol on the display (after a period of 3 hours has
passed or a defined time interval).
The user must complete his/her weighing process within this time.
After 5 minutes [
CAL 30 ] appears on the display.
This starts a “count down” of 30 seconds [
CAL 30 ] → [ CAL 0 ]. You may cancel
adjustment during these 30 seconds by pressing the TARE-key. This makes the
balance return to weighing mode in order to e. g. complete an unfinished
measurement.
After a further 5 minutes the automatic adjustment will restart and [
CAL 30 ]
appears.
You may interrupt the adjustment process several times. However, keep in mind that
an interruption of the adjustment may lead to incorrect weighing results.