50
DBS-BA-e-1311
Press ENTER to confirm selection. The currently
present switch-off criterion will be displayed.
Use navigation buttons to select the
required setting,
TIME = time-controlled
∆M = automatic
TIME setting
Press ENTER to confirm selection. The currently
set drying time is displayed.
Use navigation buttons to select the
required setting, 1 minute – 4 hours (1 minute
increments) or 4 hours - 12 hours (1 hour
increments) can be selected
Press ENTER to save.
Press ESC key to return to weighing mode. The
moisture analysis can be started (see chap. 8.2).
All the settings that have been made control the
drying process.
or
Setting
∆M
Press ENTER to confirm selection. The currently
preset value for
∆M will be displayed.
Use navigation buttons to select the
required setting,
0.01% – 0.1% (0.01% increments) can be
selected.
Press ENTER to save.
Press ESC to return to Moisture analysis mode.
The equipment is ready to measure (see chap.
8.1). All the settings that have been made
control the drying process.