DBS-BA-e-1311

• If the value of the adjustment weight needs to be changed, use the

navigation buttons to make the required setting (see chap. 7.1.1.
“Numerical input”).

• The adjustment protocol output is started upon connection to an

optional printer and activated GLP function (see chap. 6.3).

 When the adjustment weight display flashes, e.g.

50g, place it carefully in the centre of the sample
dish.







 Adjustment starts.

 Wait until „0.000g“ appears. Remove the

adjustment weight during the flashing display.

 The adjustment is completed when “END” is

displayed. The device returns to the menu
automatically.

 Press ESC to return to weighing mode.