beautypg.com

Recommended liquids – Grant Instruments TC120 Operating Manual User Manual

Page 9

background image

T100 & TC120

30423 V3

Operating Manual

Page 8

www.grantinstruments.com

EN

4.5

Removing the controller from the bridge plate


Allow the working liquid to cool before removing the T100/TC120 from the bridge plate.
Carefully remove the T100/TC120 and bridge plate together from the bath or vessel.
Take care as the pump will contain a small amount of the working liquid which will leak
out as the unit is handled. Undo the retaining nuts and remove the bridge plate. Attach
the retaining nuts to the locating threads for safe keeping.

4.6

Recommended liquids


The following table lists the recommended liquids for different temperature ranges.
Always ensure the liquid used is safe and suitable for your working temperature. If using
non-recommended heat transfer liquids, it is the responsibility of the user to conduct an
assessment to ensure the intended fluid is compatible with the T100/TC120 and vessel.
If using non-recommended heat transfer liquids it is important to set the over-temperature
cut-out to a value no higher than 25°C below the fire point of the liquid. If in doubt please
contact the Grant technical support team.

To ensure protection the overtemperature cut out (TC120 only) must be set
appropriately for the heat transfer liquid selected. See table below.

If using non-recommended heat transfer liquids it is important to set the over-
temperature cut-out to a value no higher than 25°C below the fire point of the
liquid. If in doubt please contact the Grant technical support team.

Use fume extraction when using silicone fliuids at elevated temperatures

Temp range

Recommended
liquid

Cut-out
setting

Comments

-30°C to 30°C


50% water, 50%
antifreeze (inhibited
ethylene glycol)



40°C

WARNING: Ethylene glycol is
toxic – follow the manufacturer’s
instructions.
For safe disposal consult your
local regulations.
Use a lid to reduce the dilution
of the mixture caused by
condensing water vapour from
the air, and to maintain the cool
down rate.

0°C to 30°C


80% water, 20%
antifreeze (inhibited
ethylene glycol)



40°C

5°C to 99.9°C







Water*







110°C

Water can be used but care
should be taken above 60°C as
hot vapour can be dangerous.
Use a lid or polypropylene
spheres above 60°C to ensure
good performance & reduce
evaporation.
At temperatures approaching
99°C the temperature
performance will be affected due
to localised boiling.
The units should not be used to
boil water.


This manual is related to the following products: