6.0 Technical specifications


6.1

Operating conditions

Ambient temperature range

5 to 40°C

Altitude above sea level

Up to 2,000m (6,500ft)

Operating environment

Indoor use only

Maximum relative humidity

80% RH up to 31°C decreasing to 50% RH at 40°C


6.2

Electrical details


Mains supply:

220-240V @ 50/60Hz or 110-120V @ 50/60Hz

Pollution degree:

2

Installation category:

II


Mains supply voltage fluctuations are not to exceed ±10% of the nominal supply voltage.

6.3

Product performance

T100

T100L

TC120

TC120L

Settable temperature range

0°C to 100°C

-20°C to 120°C

Min operating temperature

5°C to 100°C

-20°C to 120°C

Stability (DIN 12876)

±0.05°C *

Uniformity (DIN 12876)

±0.1°C

Max pump head pressure

-

210mBar (no flow)

Max pump flow rate

-

16l/min

Max current consumption

6.5A

12.5A

6.5A

12.5A

Heater power

1.3kW

1.4kW

1.3kW

1.4kW

* Temperature stability may be affected (±0.3C) in the presence of strong RF fields (10V/m) at
380-400MHz. This level of interference is very unlikely and is only likely to be encountered in
electrically noisy industrial locations as defined in EN61326)


6.4

Bath accessories information

ST5

ST12

ST18

ST26

ST38

Tank capacity (litres)

5

12

18

26

38

Liquid depth min/max (mm)

85/140

85/140

75/130

125/180 125/180

P5

P12

P18

Tank capacity (litres)

5

12

18

Liquid depth min/max (mm)*

85/140

85/140

85/140