beautypg.com

0 troubleshooting – Grant Instruments TC120 Operating Manual User Manual

Page 23

background image

T100 & TC120

30423 V3

Operating Manual

Page 22

www.grantinstruments.com

EN

11.0 Troubleshooting

Symptom

Possible cause

Action required

Display shows
“Cut”

Over-temperature
cut-out has operated

Check the set temperature is correct and that the
over-temperature cut-out temperature is set at least
5°C above the set temperature. Refer to section
5.3.1 for setting instructions.

If the over-temperature cut-out temperature is
correctly set but the unit still shows a “Cut” alarm
then the unit has an internal fault and must be
repaired before it is used again.

Temperature
does not rise
when expected

Set temp is lower
than liquid temp

Set temperature is
too close to ambient

Check that the bath set temperature is correct (see
section 5.2.3).

Increase the set temperature (see section 5.2.3) or fit
accessory cooling (see section 4.8).

Temp continues
to rise when not
expected

Set temp is higher
than liquid temp

Check that the bath set temperature is correct (see
section 5.2.3).

Alarm light on

Over-temperature
cut-out has operated

Low liquid level float
switch has operated

Check the over-temperature cut-out is set
appropriately (see section 5.3.1).

Check that the liquid level in the bath is adequate
(see section 6.4 for minimum fill levels).

Set temperature
too restricted

Liquid type set does
not allow required set
point

Change to different liquid type (see section 5.3.5).

Display shows
“Flot”

Liquid level has
dropped below
minimum level

Check that the liquid level in the bath is adequate
(see section 6.4 for minimum fill levels).

Display shows
“–Al-”

High temperature
warning alarm has
tripped

Check that the bath set temperature is correct (see
section 5.2.3).

Check that high temperature alarm is correct (section
5.3.4).

Check that the liquid level in the bath is adequate
(see section 6.4 for minimum fill levels).

Display shows
“Er H”

Incorrect calibration
value

The high temperature calibration point is less than
40°C above the low calibration point – choose a
higher temperature (see section 5.2.6).

Display shows
“Er L”

Incorrect calibration
value

The low temperature calibration point is less than
40°C below the high calibration point – choose a
lower temperature (see section 5.2.6).

Display shows
“Er d”

Incorrect calibration
value

The calibration value is greater than 10°C from the
set point – check thermometer reading and set point
(see section 5.2.6).

This manual is related to the following products: