beautypg.com

Using the t100 & tc120, Product description, Product indicators – Grant Instruments TC120 Operating Manual User Manual

Page 13

background image

T100 & TC120

30423 V3

Operating Manual

Page 12

www.grantinstruments.com

EN

5.2

Using the T100 & TC120

5.2.1 Product description

5.2.2 Product indicators

There are two indicator lights on the T100/TC120:

 Heater on light (orange) marked

 Alarm light (red) marked


The alarm lamp will light to indicate the following faults:

 Float switch has operated due to low liquid level

 Over-temperature cut out has operated

 Temperature probe fault


The cause of the alarm will also be shown on the display.

The display normally shows the temperature in °C of liquid being controlled.

The S(elect) button is used to select and store values whilst the F(unction) button is used
to access menu options and cancel functions.

The main dial is used to change temperature values and other settable parameters.

Float switch

Display

Alarm light

Control dial

Heater on light

Pump outlet plate
(TC120 only)

Over-temperature dial
(TC120 only)

Select button

Function button