beautypg.com

Flintec LDU 78.1 User Manual

Page 23

background image

LDU 78.1 Technical Manual, Rev. 12 May 2011

Page 23 of 36

8.6.7. GL

Get Data String “Average, Gross and Status“

Master (PC / SPS) sends

Slave (LDU XX.X) responds

Meaning

GL

L+000100+0011005109

(example)

Average value: +000100 d (no decimal point)

Gross value: +001100 d (no decimal point)

Status bit 1: 5 (not used)

Status bit 2: 1 (Hex)

Check sum: 09 (Hex)

For check sum, status bit 1 and status bit 2 see command GW.

8.6.8. OF

Output Format for Data String GW and GL

This command puts the range information and/or the decimal point into the “long” data strings of the GW and
GL output response.

Master (PC / SPS) sends

Slave (LDU XX.X) responds

Meaning

CE

E+00017

(example)

Request: TAC counter CE17

CE 17

OK

Calibration sequence active

OF 1

OK

Setup: OF = 1

Output Format

Parameter setting

Range Information

DecimalPoint in GW/GL response

0 (= factory default)

No

No

1 Yes

No

2 No

Yes

3 Yes

Yes


e.g. when the range information is selected, the data strings will change from G+00000 to Gn+00000,
where 1 ≤ n ≤ 3.