LDU 78.1 Technical Manual, Rev. 12 May 2011

Page 10 of 36

5.4. Terminal

Configuration

LDU 78.1

UA 73.2

UA 77.1

Function

Pin no.

0

Gnd (Shield)

Gnd (Shield)

Common Ground

1

+Exc

+ Exc

+ Excitation for load cell

2

+ Sens

+ Sens

+ Sense for load cell

3

+ Sig

+ Sig

+ Signal

4

– Sig

– Sig

– Signal

5

– Sen

– Sen

– Sense for load cell

6

– Exc

– Exc

– Excitation for load cell

7

Gnd

Gnd

Signal ground / 0 V DC

8 NC

NC

Not

connected

9 NC

NC

Not

connected

10

+ Rx

RxD (RS232)

Receive

11

- Rx

Gnd (RS232)

UA 73.2: Receive / UA 77.1: Common ground RS232

12

- Tx

Gnd (RS232)

UA 77.1: Senden / UA 77.1: Common ground RS232

13

+ Tx

TxD (RS232)

Transmit

14

0 In

0 In

Digital input 0 (with reference to ground)

15

0 out

0 out

Digital output 0 (with reference to ground)

16

1 In

1 In

Digital input 1 (with reference to ground)

17

1 Out

1 Out

Digital output 1 (with reference to ground)

18

+ PWR

+ PWR

Power supply 12...24 V DC

19

Gnd

Gnd

Common ground / 0 V DC

Remarks:

UA 73.2 (see chapter 5.3) is prepared for full-duplex operation
Valid for half-duplex operation: pin no. 10 + 13 = A and pin no. 11 + 12 = B

UA 77.1 with integrated RS422/RS232 converter (see chapter 5.2)


5.5. Load

Cell

Connection

The load cell wiring should be made carefully before energizing to avoid damages to the amplifier and the load
cells. The input resistance of the load cells that you want to connect should be more than 250 Ω.
The sense pins of the instrument should be connected. In 4-wire installations the switches SW1 and SW2 (see
chapter 5.1) have to be short circuited.

Flintec load cell; 4 wire connection

Flintec load cell; 6 wire connection