LDU 78.1 Technical Manual, Rev. 12 May 2011

Page 15 of 36

8.2. Calibration Commands – CE, CM n, CI, MR, DS, DP, CZ, CG, ZT, FD,

IZ, ZR, ZI, WT, TM, CS

8.2.1. CE

Read TAC* Counter / Open Calibration Sequence

With this command you can read the TAC counter (*TAC = Traceable Access Code) or you can open a
calibration sequence.

Master (PC / SPS) sends

Slave (LDU XX.X) responds

Meaning

CE

E+00017

(example)

Request: TAC counter CE17

CE 17

OK

Calibration sequence active

This command must be issued PRIOR to any attempt to set the calibration parameters CM n, CI, MR, DS, DP,
CZ, CG, ZT, ZR, FD, LC, LN n or CS. In legal for trade applications the TAC counter can be used to check if
critical parameters have been change without re-verification. After each calibration the TAC counter increases
by 1.

8.2.2. CM n Set Maximum Output Value

This command (CM n with n = 1, 2 or 3) is used to set up the maximum output value (respective the switching
point in multi range applications). Permitted values are from 1 to 99 999.

Master (PC / SPS) sends

Slave (LDU XX.X) responds

Meaning

CM 1

M+030000

Request: CM 1 = 30 000 d

CE

E+00017

(example)

Request: TAC counter CE17

CE 17

OK

Calibration sequence active

CM 1 50000

OK

Setup: CM 1 = 50 000 d

This value will determine the point at which the output will change to “oooooo”, signifying over-range respective
the point at which the output will change the measuring range / interval size.

Application

CM 1 = MAX 1

CM 2 = MAX 2

CM 3 = MAX 3

Single range

CM 1 = 1...99 999

CM 2 = 0
(means CM 2 not used)

CM 3 = 0
(means CM 3 not used)

Dual range or dual interval
( Command MR)

CM 1 = 1...MAX 1

CM 2 = MAX 1...99 999

Triple range or triple interval
Teilungen ( Befehl MR)

CM 1 = 1...MAX 1

CM 2 = MAX 1...MAX 2

CM 3 = MAX 2...99 999

It is necessary: 1  MAX 1 < MAX 2 < MAX 3  99 999

Note:

The range, in which a scale can be set to zero (SZ) or automatic zero tracking (ZT) is active, is +/- 2% of

CM value. Factory default: CM1 = 99 999, CM 2 = 0, CM 3 = 0

8.2.3. CI

Set Minimum Output Value

This command is used to set up the minimum output value. Permitted values are from – 99 999 to 0.

Master (PC / SPS) sends

Slave (LDU XX.X) responds

Meaning

CI

I

000009

Request: CI = –9 d

CE

E+00017

(example)

Request: TAC counter CE17

CE 17

OK

Calibration sequence active

CI

10000

OK

Setup: CI = –10 000 d

This value will determine the point at which the output will change to “uuuuuu”, signifying under-range.

Note:

In bipolar applications (e.g. force- or torque measurements) this parameter defines the max. output value

for input signals with negative sign. Factory default: CI =

9

8.2.4. MR Set Multi-range / Multi-interval

This command is only relevant, if CM 2 > 0 or CM 3 > 0. Is this the case,then this command defines, if the
application is multi-range or multi-interval. Permitted values are 0 (Multi-interval) or 1 (Multi-range).

Master (PC / SPS) sends

Slave (LDU XX.X) responds

Meaning

MR

M+00000

Request: MR = 0 (Multi-interval)

CE

E+00017

(example)

Request: TAC counter CE17

CE 17

OK

Calibration sequence active

MR 1

OK

Setup: MR = 1 (Multi-range)

Note:

Single range applications ignore this parameter.