beautypg.com

Buffalo Filter Porta PlumeSafe 604 User Manual

Page 9

background image

BUFFALO FILTER®

Porta PlumeSafe® 604 Smoke Evacuation System – Rev B

Page 8 of 14

FILTER LIFE INDICATOR

(Membrane Control Panel)

The filter life indicator on the membrane control panel provides a visual indication of the status of

the life of the filter in use. The filter life i ndicator for the PlumeSafe Smoke Evacuation System

will automatically adjust for the filter type used as follows:

Virosafe 6 Filter – 6 hours filter life

ViroSafe OH Filter – 6 hours filter life

ViroSafe EY Filter – 12 hours filter life

ViroSafe Vet Filter – 24 hours filter life

ViroSafe XLL Filter – 18 hours filter life

The PPS604 Filter Life Indicator is factory set. All filter life timing and reset is automatic.

Reading the Filter Life Indicator

Install an unused ViroSafe Filter into the system per the installation instructions contained in this

operator’s manual. When the sy stem is turned on, the filter l ife indicator with light up the

leftmost GREEN LED, indicating 100% filter life. The indicator will progress thru subsequent

GREEN LEDs, to an A MBER LED as time elapses and begin flashing RED to indicate t hat the

filter has expended its useful life and requires replacement.

The PPS604 filter life is factory set. When the maximum filter life is consumed and the

smoke evacuator is in standby (suction motor off) for greater than six (6) hours or the main

power is switched off – a new filter is required to activate the smoke evacuator and make

operational.

FUSES (Rear or Side Panel)

Two 10 AMP fuses (8 AMP for international s ystems) are located on the rear or side of the

system. It el ectrically protects both the sy stem and the operator from damage or injury. If the

system is overheated or if there is an electrical surge in the electrical system, fuses will break and

the system will not operate. To re-start th e system replace the fuses within the power entry

module with Buffalo Filter approved components.

When the Service light illuminates, please contact BUFFALO FILTER Technical Services

for system service instructions.

2.2 ViroSafe Filter Instructions

BUFFALO FILTER P/N

PPS604

CONFIGURATION

Portable or tabletop

ViroSafe

®

Filters -- single or double ports 7/8”

(22mm) and/or 1-

¼” (32mm)

4-Stage filtration in one casing (pre-filter, ULPA, Carbon, post-filter)

Filter(s) ULPA