1.1 Introduction

BUFFALO FILTER

®

PlumeSafe Smoke Evacuation Sy stems are designed to re move smoke and

noxious odors produced by surgical smoke during laser, electrosurgical, ultrasonic, argon, plasma and

mechanical instument procedures.

The PlumeSafe Smoke Evacuation Systems have been designed with a high suction, high fl ow rate

centrifugal action pump. The ultra-quiet motor is used to draw the surgical smoke from the surgical

site through the vacuum tubing and into the ViroSafe filter where the surgical smoke is processed by

a series of filters. A singl e disposable filter is used to simplify the installation and removal during

filter changes. The filter i s completely enclosed to protect the healthcare personnel from potential

contamination during filter changes. One Buffalo Filte r ViroSafe filter contains four different stages

within to capture the smoke plume.

The first stage filtration is a prefilter whose function is to trap and remove gross particulate and casual

fluid.

The second stage filtration is ULPA

grade (Ultra Low Penetration Air) filter whose high-tech

patented (U.S. Patent #58740 52) design captures particulates and micro-organisms from .1 to . 2

microns at an efficiency of 99.999%.

The third stage filtration uses the highest grade

virgin activated carbon, especially designed for

Buffalo Filter for the remo val and adsorption of odors and toxic g ases produced by burning tissues.

These harmful gases may constitute a health hazard to healthcare professionals who are subjected to

prolonged exposure. The activated carbon used in the Plum eSafe Smoke Evacuation Systems

preferentially removes toxic organic gases rather than water vapor and provides optimal odor

removal.

The fourth stage filtration is an expanded foam used to trap activated carbon fines from migrating out

of the filter.

The electronic controls on the face panel of th

e PlumeSafe S moke Evacuation Sy stem has been

designed “user friendly ” and facilitate unit set up and operation. Please refer to Sectio n 2.0 for

Operating Instructions.

1.2 Inspection

The PlumeSafe Smoke Evacuator has been thoroughly tested and inspected before shipment from the

factory. Please check the unit before using it to in sure that no da mage has occurred in transit. If

damage is evident, please contact BUFFALO FI LTER Customer Service at 1-800-343-2324 (U S

Only) or (716) 835-7000 (International)

In addition, please compare the accessories you receive with the standard accessories list below. If an

item is missing, please notify BUFFALO FILTER Customer Service.

Standard Accessories:

Operator’s Manual

Power Cord

Pneumatic Footswitch

System Description

Section 1.0