3.1 General Maintenance Information

This section contains information for ordinar y upkeep of the BUFFALO FILTER Plu meSafe Smoke

Evacuation System. While the system has been designed and manufactured to high industry standards, it

is recommended that periodic inspection and pe

rformance testing be perf ormed by a qualified

Biomedical Technician to ensure continued safe and effective operation.

3.2 Cleaning

The exterior of the sy stem may be cleaned by wiping it with a cloth that has been da mpened (not

dripping) with a mild detergent such as Windex® or Formula 409®. Windex® is a registered trademark

of the Drackett Products Company. Formula 409® is a registered trademark of the Clorox Corporation.

3.3 Periodic Inspection

The PlumeSafe Smoke Evacuation System should be visually inspected at least every year. This

inspection should include checks for:

Damage to the power cord.

Damage to the power plug.

Tightness of the power plug.

Proper mating, cleanliness and absence of damage to the filter inlet.

Obvious external or internal damage to the system.

3.4 Troubleshooting the System – see below.

PROBLEM

POTENTIAL CAUSE

CORRECTIVE ACTION

1. Smoke Evacuation

System is ON but suction is

minimal or none

1. Filter not seated

completely

2. Filter clogged
3. Vacuum hose clogged
4. Motor/blower obstructed

1.

Re-install ViroSafe Filter, press firmly into

place

2.

Replace filter – use ViroSafe EY filter and

pre-filters for Erbium:YAG laser procedures

3.

Check or replace Smoke Evacuation Tubing.

4.

Call BioMed or BUFFALO FILTER

Technical Services at: 1-800-343-2324 or

(716) 835-7000

2. Smoke Evacuation

System does not function

even though suction ON

button is depressed

1. Not plugged into

electrical outlet

2. Fuses blown
3. Footswitch plugged in
4. Electronic system failure
5. Main power switch in

OFF position

6. Filter life expired or

invalid filter

1. Check power outlet and connection to rear or

side panel of machine

2. Replace fuses located within the power entry

module.

3. Unplug footswitch
4./5.Check main power switch on rear or side of

system to ensure that it is in on the ON (|)

position

6. Call BioMed or BUFFALO FILTER

Technical Services at: 1-800-343-2324 or

(716) 835-7000

Maintenance

Section 3.0