beautypg.com

Mounting and manual measurements – ALTANA BYK-mac i ROBOTIC Manual User Manual

Page 12

background image

11

Mounting and Manual Measurements

3. Mounting and Manual Measurements

Before operating the instrument for the first time,

please read the operating manual and take

particular notice of the Safety Instructions.

3.1 Mountig of instrument

The measuring instrument is fixed on the mounting

board with three screws (arrow).
The millings on the fixing bar and a nib ensure

proper attachment.
A drawing, that contains all dimensions for

producing the mounting board, is enclosed with

each instrument. The drawing also describes the

measurement plane.
In order to be able to do the fine positioning of the

instrument, a prepositioning to < ± 2 mm distance

to the measurement plane is necessary.
The 4 pins measure the distance to the surface and

pass this value as well as the tilting and transverse

angle on to the robotic control.
Attention must be paid to the measurement

direction (see illumination direction on instrument

housing).