P205 will appear in the “Custom Setup” list for each receive port enabled.
For example, if the receive mode is enabled for all four ports, the list will
display
P205—
,
P205˜
,
P205™
, and
P205š
.
P240 – Minimum Transmit Width
A weight value that cannot be displayed due to the 6-digit limitation of the
standard VF display may not be printed. To ensure this is not possible, P240
must be set to a width of not greater than 7 (6 digits and a decimal point).
NTEP must also be enabled at P440. Any weight value that exceeds the
minimum width specified will be printed as dashes "-------".
P440 – NTEP Enable
P440 appears in the “Custom Setup” list if disabled. Refer to Enabling
NTEP Operation on page 46 for possible implications.
P9990 – Macro Instance Selection
P9990 appears in the “Custom Setup” list if at least one macro is configured.
Macro operation should be checked to verify its conformance to all
regulations.
A
DDITIONAL
C
ONFORMANCE
C
ONSIDERATIONS
Several parameters must be considered on an individual basis as their
configuration may vary with different applications. These parameters
include, but are not limited to those listed in Table 5.
Table 5: Additional Conformance Parameters
P
ARAMETER
D
ESCRIPTION
C
OMMENT
P110
Full Scale Capacity
Verify proper scale capacity.
P111
Division Size
Verify allowable scale divisions.
P112
Zero Track
Verify required selection.
P114
Motion
Verify required selection.
P118 Zero
Range
Verify required selection.
P212
Print Stability
Verify required selection.
P126 Æ P130
Multi-Range
Verify proper configuration.
P151 Æ P154
Custom Units
Verify name and conversion factor.
P600 Æ P646
Rename Parameters
Verify acceptable names.