Pecifications – Holtgreven GSE-665 User Manual
Page 19
C
H A P T E R
1 : I
N T R O D U C T I O N
11
S
PECIFICATIONS
Performance
• Full Scale
Selectable
• Resolution
100,000 displayed (+/-500,000 internal)
• Display Update
Selectable 0.05-20 seconds
• A/D Conversion
60 Hz
• Non-Linearity
0.005% of full scale (input dependent)
• Zero Track
0.05-20.0 displayed divisions
• Zero Range
Selectable from 0.01-100% of full scale
• Calibration
Selectable, 5 multi-point calibration for
linearization
• Division Size
.00001 through 500
• Warranty
2
year
Electrical
• Power input 90-250 VAC, 50/60 Hz; 10-32 VDC
• Fuse 0.8 amp time lag
Load Cell Input
• Connections 4 lead or 6 lead with sense, jumper selectable
• Power 14 350-ohm cells (560 and 660 Series)
• Power 12 350-ohm cells (460 Series)
• Signal Range 0.1 to 20 mV/V at full scale
• Excitation 10 VDC, short circuit protected
• Current 400 mA maximum
Communications
• Bi-directional RS-232 serial ports
• Independent custom transmit tables for labels, reports, transmitting
control characters, etc.
• Transmit output: continuous transmit, on request, motion inhibited
• Selectable baud rate
• Selectable protocol