9. Key in the full scale capacity for the load cell(s) and press [ENTER].

For the Model 460, use the [PRINT] and [UNITS] keys to enter
numeric values. Refer to Table 2: 460 Key Functions on page 8.
- or -
Press [ENTER] to accept the displayed value.

10. The display briefly shows Updtg Gains as it updates the gain values,

then prompts CurWt Zero?.

11. Press [ENTER] to establish the current input signal as the zero

reference.
- or -
Press [SELECT] to display Zero=0mVv?. Press [ENTER] to use a
0mV/V output as the zero reference.
- or -
Press [SELECT] to display Keyin CurWt. Key in the known gross
weight already applied to the scale and press [ENTER].
- or -
Press [CLR] to bypass the zeroing option.

12. The display shows CAL OK? suggesting that the calibration is

acceptable.

Accept the calibration by pressing [ENTER] at the CAL OK? prompt.
- or -
Repeat the calibration by pressing [CLR] at the CAL OK? prompt.

13. Once the calibration is accepted, press [ENTER] at the ENTER=SAVE

prompt and again at the ENTER=EXIT prompt to save the new
calibration and exit the calibration mode.
- or -
To exit the calibration mode without saving the new calibration, press
[CLR] at the ENTER=SAVE prompt. Then press [ENTER] at the
ENTER=UNDO

prompt and again at the ENTER=EXIT prompt to

exit the calibration mode.

S

AVE

C

ALIBRATION

The indicator will perform the calibration, display the value of the
calibration weight, and display the prompt, CAL OK?. At this point, you can