beautypg.com

Holtgreven GSE-665 User Manual

Page 35

background image

C

H A P T E R

3 : C

A L I B R A T I O N

27

1. Remove all weight from the scale.

2. Access the Calibration Mode (see page 23).

3. Press

[SELECT] to toggle to the Only Zero? routine.

4. At

the

Only Zero? prompt, press [ENTER].

The displayed value zeroes out. Adj Zero! is displayed briefly, followed
by CAL OK?.

5. Press

[ENTER] to accept the newly established zero, or [CLR] to

repeat the calibration.

9. At

the

Enter = Save prompt, press [ENTER]

10. To return to the Weighing Mode, at the ENTER = EXIT prompt, press

[ENTER].

C

AL

R

ESET

Cal Reset may be necessary when an over-load or under-load condition
exists, preventing the completion of the calibration process. Calibration
Reset adjusts the zero and gain factors of the A/D amplifier to factory
default values for maximum sensitivity.

After performing a calibration reset, a complete recalibration is required.
The effects of a calibration reset do not take effect until the indicator is
recalibrated and calibration information has been saved.

If Code 02 (under-load) or Code 03 (over-load) is displayed during
calibration, press [CLR] to perform a calibration reset.


When reset, these parameters are adjusted to the lowest possible values.
Contact your GSE Distributor regarding Information Parameters for a more
detailed explanation on the gain factors.

Normally, a Cal Reset is performed if the amplifier is locked in at an
extremely high gain factor and will not allow a new calibration to be
performed.

1. Access the Calibration Mode (see page 23).

2. Press

[SELECT] to toggle to the Cal Reset.

This manual is related to the following products: