Using the device – FRITSCH PULVERISETTE 14 User Manual
Page 38
Carry out the installation and removal of the grinding elements as well as
the milling process as described in Ä Chapter 6.3 ‘Installing the grinding
elements’ on page 27, Ä Chapter 6.5 ‘Removing the grinding elements’
on page 29 and Ä Chapter 6.4 ‘Conducting a grinding operation’
on page 28. The only difference is that the tube to the collecting vessel is
inserted in and fastened to the bent tube support.
1.
Position all devices as shown and described above, then connect
to the mains.
There are two plugs on the exhaust system: a mains
plug and an earthing plug. Plug both into the socket.
2.
Prepare and close the PULVERISETTE 14 as described in
Chapter 6.3 ‘Installing the grinding elements’ on page 27.
3.
Slide the rubber sleeve with the exhaust hose from the cyclone
separator over the outlet of the grinding vessel.
4.
Fasten the other end of the exhaust hose on the open hose con-
necting tube of the cyclone separator.
There are 2 opposing hose connections on the cyclone
separator. Do not use the sealed connection; leave it
closed. This would increase the noise level!
5.
Screw the sample bottle onto the adapter on the bottom of the
cyclone separator; make sure the exhaust hose is firmly connected
and switch on the exhaust system.
6.
Switch on the PULVERISETTE 14.
Installing and removing the grinding ele-
ments
Working with the sample exhaust
system
Using the device
- 38 -