Checkline RH6 User Manual
Page 10
Electromatic Equipment Co., Inc.
-10-
5. Now put the third gasket ring into the bottom part.
6. Fit the metal ring on the third gasket.
7. Take the upper part with the RH6 and attach these carefully at the bottom part of
the calibration device.
8. Pick up the RH6 together with the calibration device STRAIGHT and DON´T
TURN IT AROUND. Screw it up like shown in the picture.
Then put the RH6 with the calibration device down on a table carefully and proceed
as follows:
Conditioning the sensor
To achieve best results, let the sensor adjust for two hours. The temperature has
to be between 20°C and 26°C.
If the shown measuring value differs more than the factory tolerance (1.5% r.h.), we
recommend to carry out a recalibration as follows.
Electromatic Equipment Co., Inc.
-11-
11.0 OFFSET ADJUSTMENT
1. Leave the measuring device in the calibration equipment, and switch it on.
2. Press the Rolling Menu button until you reach the main menu.
3. Select the menu item Options by pressing the button and confirm by pressing
.
4. Navigate to
Setting
using the button and confirm by pressing
again.
5. Enter the superuser password using the buttons
resp.
and confirm by
pressing .
•
The superuser password after consignment is the serial number of the
device, shown on the display after switching on the device or in menu item
Status
.
6. A query if a setting is desired appears. Confirm by
pressing .
7. Wait until the bar has risen completely. The device
adjusts by itself and automatically jumps back to the
measuring window. The adjustment is completed
now.
8. Check the result before you remove the device from
the calibration equipment. Depending on the
temperature the display should show a humidity
around 50% now.
If you made a mistake during the setting, you can reset to the factory calibration as
follows: