beautypg.com

Torque ranges – Checkline CAP-TNP User Manual

Page 2

background image

- 2 -

1.0 I

NTRODUCTION

Thank you for choosing the CAP-TNP digital torque tester, with proper care this
unit will provide many years of reliable service.

The CAP-TNP torque meter is a portable device that can be used as a quality
assurance tool for various applications that requires turning (both opening and

closing) and twisting.

Some fields that can utilize this equipment are:

• Pharmaceutical,
• Food and Beverage
• Cosmetic products

Built with internal rechargeable batteries, the CAP-TNP can operate as a portable
DC device or thru the universal AC adapter. Designed with a small footprint it can
easily be moved around the shop floor or the laboratory to maximize use.

The programmable HI-LO set points make this unit ideal for pass-fail testing in a
production environment.

1.1 Complete

Kit

Each package includes the following:

• Operations Manual
• Software Manual
• Software Installation Disk

(Digitorq Software)

• Warranty Cards
• USB A to USB B communication cable
• Universal AC adapter (100-240 VAC)
• Set of 30 mm chuck pins (4 pieces)

IMPORTANT: Upon receiving the unit, please check for any obvious physical
damage that may have occured during shipping. If any damage is found, please
notify your carrier immediately before shipping the unit back to Electromatic for
repairs and inspections.

WARNING

Do not test products that are filled with liquid contents. The CAP-TNP
torque tester is not protected from liquid spills that may come from the
tested product.

- 15-

5.0 S

PECIFICATIONS

OverloadProtection

150%

Sample Diameter Range (min to max) 0.39" to 7.48" (10–190 mm)

Overload display

Display "OVR" on LCD (blinking on/off)

Main display

4-digit LCD display

Character height 12mm

Sub display

3-digit LCD display

Character height 7mm

Comparator display

Hi, GO, Lo LED Indicators

Accuracy

± 0.5% full scale

Open mode

Max vale when opening.
Displays max counter clockwise torque.

Close mode

Max value when closing.
Displays max clockwise torque

Average mode

Real time display. Displays max torque in real time

Display Update

1, 2, 4 or 8 updates/second, user-set

Sampling Rate

1000 times/second

Memory Storage

1000 data points (max)

Statistic process

Average value, max value and min value

Data output

USB 1.1

PC software

Digitorque software

Power

Built in nickle hydride battery or Auto-ranging AC
adapter (AC 100 — 240V)

Dimensions

12.60" x 8.94" x 4.96" (320 x 227 x 126mm) L x W x H

Operation Temp.

32 - 104°F (0 - 40°C)

Battery Life

8 hours after full charge

Battery Recharge Time

Max. 16 hours

Battery Type

NiMH

CAP-TNP-2

0-2.000 Nm
0-200.0 Ncm
0-20.39 Kgcm

0-17.70 Lbin

CAP-TNP-5

0-5.000 Nm
0-500.0 Ncm
0-50.99 Kgcm
0-44.25 Lbin

CAP-TNP-10

0-10.00 Nm
0-1000 Ncm
0-102.0 Kgcm
0-88.50 Lbin

Model Capacity Resolution

0.001 Nm
0.1 Ncm
0.01 Kgcm
0.01 Lbin

0.001 Nm
0.1 Ncm
0.01 Kgcm
0.01 Lbin

0.01 Nm
1 Ncm
0.1 Kgcm
0.1 Lbin

Torque Ranges