Series TT02 Digital Torque Tool Testers
User’s Guide
21
10. After all the clockwise calibration points have been completed, the display will appear as follows:
Press ENTER.
11. At the completion of the counter-clockwise calibration, the display will appear as follows:
To save the calibration information, select “SAVE & EXIT”. To exit without saving the data select
“EXIT W/O SAVING”.
12. Any errors are reported by the following screens:
Displayed at the start of calibration if a disallowed unit is selected.
Ensure that the load is not swinging, oscillating, or vibrating in any manner. Then try again.
The calibration load does not match the set value.
LOAD NOT STABLE
PLEASE TRY AGAIN
CALIBRATION
Units must be kgFmm.
PLEASE TRY AGAIN
PRESS ENTER
CALIBRATION
CLOCKWISE
LOAD TOO LOW
PLEASE TRY AGAIN
CALIBRATION
COMPLETE
SAVE & EXIT
EXIT W/O SAVING
CALIBRATION
CLOCKWISE COMPLETE
REVERSE DIRECTION
FOR CCW
Attach necessary
weight fixtures.
THEN PRESS ENTER