Checkline TT02 User Manual
Page 20
Series TT02 Digital Torque Tool Testers
User’s Guide
19
example, Model MTT02-50 (with capacity of 50 lbFin) should be calibrated at 10, 20, 30, 40, and
50 lbFin loads in each direction.
2. To escape the Calibration menu at any time, press ESCAPE. The display will appear as follows:
Selecting “CANCEL” will revert back to the Calibration setup. Selecting “EXIT W/O SAVING” will
return to the menu without saving changes.
3. After the number of calibration points has been entered, press ENTER. The display will appear as
follows:
4. Place the TT02 in a fixture free from vibration, then press ZERO. The tester will calculate offsets,
and the display will appear as follows:
If failed:
5. The following screen appears after the offsets have been calculated:
CALIBRATION
NOT COMPLETE
CANCEL
EXIT W/O SAVING
CALIBRATION
CLOCKWISE
Attach necessary
weight fixtures.
THEN PRESS ENTER
CALIBRATION
OFFSET
Sen.Offset Adj.Failed
Ana.Offset Adj.Failed
CALIBRATION
OFFSET
Sen.Offset Adj.Passed
Ana.Offset Adj.Passed
CALIBRATION
OFFSET
Please wait…
CALIBRATION
OFFSET
Place tester
vertical
THEN PRESS ZERO