Series TT02 Digital Torque Tool Testers

User’s Guide


19

example, Model MTT02-50 (with capacity of 50 lbFin) should be calibrated at 10, 20, 30, 40, and
50 lbFin loads in each direction.

2. To escape the Calibration menu at any time, press ESCAPE. The display will appear as follows:

Selecting “CANCEL” will revert back to the Calibration setup. Selecting “EXIT W/O SAVING” will
return to the menu without saving changes.

3. After the number of calibration points has been entered, press ENTER. The display will appear as

follows:

4. Place the TT02 in a fixture free from vibration, then press ZERO. The tester will calculate offsets,

and the display will appear as follows:

If failed:

5. The following screen appears after the offsets have been calculated:

CALIBRATION

NOT COMPLETE


CANCEL

EXIT W/O SAVING

CALIBRATION

CLOCKWISE

Attach necessary
weight fixtures.

THEN PRESS ENTER

CALIBRATION

OFFSET


Sen.Offset Adj.Failed
Ana.Offset Adj.Failed

CALIBRATION

OFFSET

Sen.Offset Adj.Passed
Ana.Offset Adj.Passed

CALIBRATION

OFFSET


Please wait…

CALIBRATION

OFFSET

Place tester
vertical
THEN PRESS ZERO