Series TT02 Digital Torque Tool Testers

User’s Guide


18


11.3 Auto Output Settings
Select the output type. Select RS-232/USB and/or Mitutoyo outputs, and select First and/or Second
peaks. The display will appear as follows:



13 CALIBRATION

13.1 Initial Physical Setup
The TT02 should be mounted to a fixture rugged enough to withstand a torque equal to the full capacity of
the instrument. Suitable certified calibration equipment is required, and caution should be taken while
handling such equipment.

13.2 Calibration Procedure

1. Select

Calibration from the menu. The display will appear as follows:

The tester can be calibrated at up to 10 points in each direction. Enter the number of calibration
points for each direction. At least one point must be selected for each direction.

Note: To achieve the accuracy specification of ±0.3% of full scale, it is recommended to calibrate
the TT02 at 5 or more even intervals in both the clockwise and counter-clockwise directions. For

AUTO OUTPUT

SETTINGS


RS232/USB Output

Mitutoyo Output
First Peak
Second Peak

CALIBRATION

ENTER # CAL POINTS

(1 TO 10)

CLOCKWISE:

5

COUNTER-CLOCKWISE:

5