13 CALIBRATION
13.1 Initial Physical Setup
The TT02 should be mounted to a fixture rugged enough to withstand a torque equal to the full capacity of
the instrument. Suitable certified calibration equipment is required, and caution should be taken while
handling such equipment.
13.2 Calibration Procedure
1. Select
Calibration from the menu. The display will appear as follows:
The tester can be calibrated at up to 10 points in each direction. Enter the number of calibration
points for each direction. At least one point must be selected for each direction.
Note: To achieve the accuracy specification of ±0.3% of full scale, it is recommended to calibrate
the TT02 at 5 or more even intervals in both the clockwise and counter-clockwise directions. For
AUTO OUTPUT
SETTINGS
RS232/USB Output
Mitutoyo Output
First Peak
Second Peak
CALIBRATION
ENTER # CAL POINTS
(1 TO 10)
CLOCKWISE:
5
COUNTER-CLOCKWISE:
5