Checkline TT02 User Manual
Page 18
Series TT02 Digital Torque Tool Testers
User’s Guide
17
Any combination of the above functions may be selected.
12.2 Settings
The display will appear as follows:
Thresholds and percentage drops are illustrated below:
Function Description
Enabled
If enabled, 2PK will appear as one of the operating modes. In the main display,
the Peak readings will reference the first and second peaks – first peak on top,
second peak below. Refer to the Home Screen & Control section for details.
Peak Settings
Press ENTER to access the Peak Settings sub-menu. See Section 12.2 for
details.
Auto Output
Press ENTER to access the Auto Output Settings sub-menu. See Section 12.3
for details.
Auto Store PK1
Automatically stores the first peak reading to memory.
Auto Store PK2
Automatically stores the second peak reading to memory.
Auto Zero
Automatically zeroes the display following data transmission and/or storage.
Threshold 1
Sets the percentage of full scale at which the first/second peak detection
feature becomes active. This threshold is provided to ignore peaks that can
occur during sample loading and unloading. Available settings: 1-90%, in
1% increments between 1-5%, and in 5% increments between 5-90%.
Percentage Drop 1 Identifies the first peak via detection of a specified percentage drop from
peak.
Threshold 2
Same as with Threshold 1, but refers to a percentage of full scale beyond
the first peak. For example, for a 50 lbFin capacity tester, if the first peak is
20 lbFin, and Threshold 2 is set to 15%, the threshold is equal to 27.5 lbFin.
Percentage Drop 2 Same as Percentage Drop 1, for the second peak.
Auto Zero Delay
Sets the time delay before the primary and peak readings are zeroed.
Available settings: 1-60 sec., in 1 sec. increments between 1-5 sec., and in
5 sec. increments between 5-60 sec.
PEAK SETTINGS
Thresh. 1:
5 %
% Drop 1: 10 %
Thresh. 2: 5 %
% Drop 2: 10 %
Auto Zero Delay
3 sec.
FIRST/SECOND PEAK
*
Enabled
+ Peak Settings
+ Auto Output
* Auto Store PK1
Auto Store PK2
* Auto Zero