Checkline WTTM-GEN1 User Manual

Page 5

background image

– 5 –

5.0 O

VERVIEW

0

1. ON/OFF

Button

Switches the instrument on/off

0

2. RESET

Button

For zero-setting after inserting the test sample

0

3. PEAK Button

For selecting the display function, current measurement
result or peak value of the tractive force (drag indicator
function), and for calling up the peak value of the
measurement

0

4. MODE

Button

For selecting the desired measurement unit

0

5. Display

0

6. Outlet

For AC Adapter

0

7. Outlet

For data transfer

0

8. Terminal

adapter with 12 slots for inserting the cable joint

0

9. Tractive force fixture for inserting the free cable end (with protection cap)

10. Selector switch for pull speed (access protected by screw connection)

5.1 Operating elements

– 16 –

10.0 T

ROUBLESHOOTING

No indication on display

Plug in the AC-adapter and check the electrical
connections.

RS-232C communication Check the wiring of the connection cable and the
is disturbed

compliance of wiring within the input terminals
of the PC-card.

Error Code Indications:

OV+

Overload of +120%. Reduce the load until the
instrument indicates a value within the admissible

measuring range.

OV-

Overload of –120%. Reduce the load until the

instrument indicates a value within the admissible

measuring range.

OVP

Overload +!! The load cell can be damaged. Remove the
excessive load immediately and check the instrument. If
the indicated measuring values are obviously incorrect,
the load cell must be replaced by the manufacturer.

OVM

Overload -!! The load cell can be damaged. Remove the
excessive load immediately and check the instrument. If
the indicated measuring values are obviously incorrect,
the load cell must be replaced by the manufacturer.

ERR –3-

Reading error of the E_proms. Switch off the instrument
and turn it on again. Should the error still exist, the
microprocessor may need to be replaced by the
manufacturer.

ERR –4-

Reading error of the E_proms. Switch off the instrument
and turn it on again. Should the error still exist, the
microprocessor may need to be replaced by the
manufacturer.

Pull force fixture

Remove the AC-adapter/charger, close the protection

does not work

cover and wait for approx. 5 minutes. Plug in the
AC-adapter again and switch on the instrument. If the
display does not show any of the error codes above and
still does not work, please send the instrument back for
inspection to your dealer or directly to the
manufacturer.