beautypg.com

Cannon Instrument TESC User Manual

Page 29

background image

TESC Operator's Manual, Rev 1.0

23

CANNON®

Calibrating the TESC System

e. In the Parameter Menus, open

EZ-ZONE PM > Setup > Analog Input

. The

Calibration Offset is shown in the display window.

10. When the DV2T test completes, save the test data from the DV2T to a CSV file on a USB

drive and transfer it to the PC. On the PC, use a spreadsheet to open the CSV file. In the
sample information, add the sample ID.

Note:

Because the TESC System is calibrated during manufacturing,

only small adjustments of ± .1 °C should be necessary to fine tune the

temperature setting.

If desired, the actual temperature of the sample may be determined

using a viscosity calculation tool, such as VISDISK®.

𝑛𝑛𝑛𝑛𝑛𝑛 𝑜𝑜𝑜𝑜𝑜𝑜𝑜𝑜𝑛𝑛𝑜𝑜 = 𝑐𝑐𝑐𝑐𝑐𝑐𝑐𝑐𝑛𝑛𝑛𝑛𝑜𝑜 𝑜𝑜𝑜𝑜𝑜𝑜𝑜𝑜𝑛𝑛𝑜𝑜 + �

𝑎𝑎𝑐𝑐𝑜𝑜𝑐𝑐𝑎𝑎𝑎𝑎 𝑣𝑣𝑣𝑣𝑜𝑜𝑐𝑐𝑜𝑜𝑜𝑜𝑣𝑣𝑜𝑜𝑣𝑣 − 𝑜𝑜𝑛𝑛𝑜𝑜𝑜𝑜𝑛𝑛𝑡𝑡 𝑣𝑣𝑣𝑣𝑜𝑜𝑐𝑐𝑜𝑜𝑜𝑜𝑣𝑣𝑜𝑜𝑣𝑣

𝑅𝑅𝑎𝑎𝑜𝑜𝑛𝑛 × 10

Where:

Rate of viscosity response to temperature change:

600 mPa(s) per 0.1 °C for N27C at –40 °C

2200 mPa(s) per 0.1 °C for N480B at –26 °C

11. Using the data in the spreadsheet, compare the measured viscosity at the highest torque

reading to the calibrated viscosity of the sample.

If the measured viscosity is higher than the calibrated viscosity, then the

temperature of the sample is lower than the target temperature. Subtract 0.1 °C from
the value shown for the Calibration Offset and enter this new value into the
Watlow® configurator program. Continue to step 11.

If the measured viscosity is lower than the calibrated viscosity, then the temperature

of the sample is higher than the target temperature. Add 0.1 °C to the value shown
for the Calibration Offset and enter this new value into the Calibration configurator
program. Continue to step 11.

If the measured viscosity is within the tolerance range of the calibrated viscosity of

the sample, the temperature of the sample is correct. Calibration is complete, go to
step 16.