Run quick verification test, Table 7 — tools and materials required – Cannon Instrument TESC User Manual

Page 18

background image

TESC Operator's Manual, Rev 1.0

12

Getting Started

CANNON®

Run quick verification test

This test verifies system calibration and provides a technician with basic training in operating
the CANNON® Thermoelectric Sample Conditioner System.

Table 7 — Tools and materials required

Description

Description

Assembled TESC System with mounted
Brookfield® viscometer

20 mL of an appropriate reference sample

25 mm × 150 mm rimless, disposable test
tube

Brookfield® #4B2 insulated spindle

Sample preparation materials

USB flash drive with program files

1. Turn on DV2T and TESC. If the Watlow® controller starts to run a program, press the red

EZ

button to halt the test. (Steps symbol

indicates a running program.)

2. Autozero DV2T:

a. Remove spindle if present.
b. Lower viscometer to measurement position.
c. Level TESC System if necessary.
d. Select

Next

on the DV2T

AutoZero

screen.

e. When the DV2T indicates that the AutoZero is complete, raise viscometer.

3. Prepare a 20 mL sample according to your laboratory practices:

a. Insert a 25 mm × 150 mm test tube containing 20 mL (± 0.2 mL) of the reference

sample into the test chamber and gently hand-tighten the cone ring.

b. Attach a clean #4B2 spindle to the viscometer.
c. Carefully lower viscometer into measurement position.

4. Set the TESC System temperature:

a. Using the control buttons on the Watlow® controller, set the target temperature,

which depends on the characteristics of the reference sample you are testing.

b. The TESC System immediately begins cooling or heating to achieve the target

temperature.