Table 9 — thermal conditioning programs – Cannon Instrument TESC User Manual

Page 22

background image

TESC Operator's Manual, Rev 1.0

16

Running an ASTM D2983 Test

CANNON®

Table 9 — Thermal conditioning programs

Program Name

Test
Temperature

Preheat
Temperature

Typical Product

D2983-TESC-Minus12-GO-
WatlowPgm.wcf

–12 °C

90 °C

Gear oils

D2983-TESC-Minus26-GO-
WatlowPgm.wcf

–26 °C

90 °C

Gear oils

D2983-TESC-Minus40-GO-
WatlowPgm.wcf

–40 °C

50 °C

Gear oils

D2983-TESC-Minus40-ATF-
WatlowPgm.wcf

–40 °C

90 °C

ATF, hydraulic, and
similar fluids

g. Click

Exit

.

Note:

The Watlow® calibration offset is unique at each test

temperature; the offset for –40 °C may not be the same offset for –26 °C.

3. Using the Watlow® configurator program, verify that the proper calibration offset has

been entered into the program.

4. Autozero DV2T:

a. Remove spindle if present.
b. Lower viscometer to measurement position.
c. Level TESC System if necessary.
d. Press

Next

on the DV2T

AutoZero

screen.

e. When the DV2T indicates AutoZero complete, raise viscometer.

5. Prepare a 20 mL sample according to your laboratory practices:

a. Insert a 25 mm × 150 mm test tube containing 20 mL ± 0.2 mL of the reference

sample into the test chamber and gently hand-tighten the cone ring.

b. Attach a clean #4B2 spindle to the viscometer.
c. Carefully lower viscometer into measurement position.