beautypg.com

Calibration procedure – Cannon Instrument TESC User Manual

Page 27

background image

TESC Operator's Manual, Rev 1.0

21

CANNON®

Calibrating the TESC System

Calibration procedure

Use the following procedure to calibrate the Watlow® controller.

1. Turn on DV2T and TESC. If the Watlow® controller starts to run a program, press the

red

EZ

button to halt the test. (Steps symbol indicates a running program.)

Note:

The 8 hour TESC calibration program lowers the temperature of

the sample to the desired soak temperature and maintains that

temperature for 8 hours, allowing the operator to run multiple

calibration tests.

2. Load the appropriate 8 hour TESC temperature calibration program into the Watlow®

controller:

a. Connect PC to Watlow® controller using RS-485 to USB cable.
b. Launch Watlow® configurator program:

Start Menu > EZ-ZONE Configurator > EZ-

ZONE Configurator

.

c. Select

Download a configuration file to a device

. Click

Next

.

d. Select the COM port to which the Watlow® controller is connected. Click

Next

.

e. Select

Load Program from File

.

f. Select program:

For –26 °C, select file

D2983-TESC-Caliboffset-8h-Minus26-WatlowPgm.wcf

.

For –40 °C, select file

D2983-TESC-Caliboffset-8h-Minus40-WatlowPgm.wcf

.

g. Click

Exit

.

3. Autozero DV2T:

a. Remove spindle if present.
b. Lower viscometer to measurement position.
c. Level entire unit if necessary.
d. Select

Next

on the DV2T

AutoZero

screen.

e. When the DV2T indicates AutoZero complete, raise viscometer.