I. INTRODUCTION

The Brookfield DV1 Viscometer series has provided exceptional value for viscometer users since its

introduction in 1981. Brookfield has continued to develop and improve the DV1 to maintain its position

in the market as the best value for Q/C applications. The Brookfield DV1 Viscometer continues in this

tradition of innovation, quality and value. The incorporation of a large graphical display and “Hot Keys”

has allowed for a new and improved user interface that preserves single speed data collection methods

(or multiple speeds when using Wingather SQ), which makes using the DV1 Viscometer easier than

ever before.
The Brookfield DV1 Viscometer measures fluid viscosity which is a measure of a fluid’s resistance to

flow. You will find a detailed description of the science of viscosity in the Brookfield publication “More

Solutions to Sticky Problems”, a copy of which was included with your DV1.
The principle of operation of the DV1 is to drive a spindle (which is immersed in the test fluid) through

a calibrated spring. The viscous drag of the fluid against the spindle is measured by the spring deflection.

Spring deflection is measured with a rotary transducer. This system provides continuous sensing and

display of the measurement during the entire test. The measurement range of a DV1 (in centipoise

or milliPascal-seconds) is determined by the rotational speed of the spindle, the size and shape of the

spindle, the container the spindle is rotating in, and the full-scale torque of the calibrated spring.
There are four basic spring torque series offered by Brookfield:

Spring Torque

Model

dyne•cm

milli Newton•m

DV1MLV

673.7 0.0673

DV1MRV

7,187.0 0.7187

DV1MHA

14,374.0 1.4374

DV1MHB

57,496.0 5.7496

The higher the torque calibration, the higher the measurement range. The viscosity measurement range

for each torque calibration may be found in Appendix B.

The DV1 is configured to accept an optional temperature probe, which allows temperature readout over

the range -100°C to +300°C (-148°F to + 572°F). This option allows ambient temperature measurement

or temperature measurement of the sample during viscosity testing. Contact Brookfield or your local

Brookfield dealer for more information on this instrument option.

All units of measurement are displayed according to either the CGS system or the SI system.

1. Viscosity appears in units of centipoise (shown as “cP”) or milliPascal-seconds (shown as

mPa•s) on the DV1 Viscometer display.

2. Torque appears in units of dyne-centimeters or Newton-meters (shown as percent “%”)

in both cases) on the DV1 Viscometer display.

Tp

3. Temperature appears in units of Celsius (shown as C) or Fahrenheit (shown as F) on the

DV1 Viscometer display.

The following applies to DV1 Viscometers when using an optional temperature probe. Look for the

symbol

Tp

throughout this manual for instructions pertaining specifically to DV1 Viscometers with

temperature probe option.