VISCOMETER MODEL

TORQUE

CONSTANT TK

MODEL CODE ON

DV1 SCREEN

DV1MLV

0.09373

LV

2.5xDV1MLV

0.2343

2.5LV

5xDV1MLV

0.4686

5LV

1/4 DV1MRV

0.25

1/4RV

1/2 DV1MRV

0.5

1/2RV

DV1MRV

1

RV

DV1MHA

2

HA

2xDV1MHA

4

2HA

2.5xDV1MHA

5

2.5HA

DV1MHB

8

HB

2xDV1MHB

16

2HB

2.5xDV1HB

20

2.5HB

Table D-2

The full scale viscosity range for any DV1 model and spindle may be calculated using the equation:

Full Scale Viscosity Range [cP] = TK * SMC * 10,000

RPM

where:

TK = DV1 Torque Constant from Table D-2

SMC = Spindle Multiplier Constant from Table D-1
The Shear Rate calculation is:
Shear Rate (

1

/

Sec

) = SRC * RPM

where: