Brookfield DV1 Viscometer User Manual
Page 44

Brookfield Engineering Laboratories, Inc.
Page 44
Manual No. M14-023
VISCOMETER MODEL
TORQUE
CONSTANT TK
MODEL CODE ON
DV1 SCREEN
DV1MLV
0.09373
LV
2.5xDV1MLV
0.2343
2.5LV
5xDV1MLV
0.4686
5LV
1/4 DV1MRV
0.25
1/4RV
1/2 DV1MRV
0.5
1/2RV
DV1MRV
1
RV
DV1MHA
2
HA
2xDV1MHA
4
2HA
2.5xDV1MHA
5
2.5HA
DV1MHB
8
HB
2xDV1MHB
16
2HB
2.5xDV1HB
20
2.5HB
Table D-2
The full scale viscosity range for any DV1 model and spindle may be calculated using the equation:
Full Scale Viscosity Range [cP] = TK * SMC * 10,000
RPM
where:
TK = DV1 Torque Constant from Table D-2
SMC = Spindle Multiplier Constant from Table D-1
The Shear Rate calculation is:
Shear Rate (
1
/
Sec
) = SRC * RPM
where:
SRC = Shear Rate Constant from Table D-1