Small sample adapter and thermosel, Ul adapter – Brookfield DV1 Viscometer User Manual
Page 37
Brookfield Engineering Laboratories, Inc.
Page 37
Manual No. M14-023
Small Sample Adapter and Thermosel
SSA and
Thermosel
Spindle
Viscosity Range (cP)
Shear Rate sec-1
DV1MLV
SC4-16
120 - 400,000
.29N
SV4-18
3 -
10,000
1.32N
SC4-25
480 - 1,600,000
.22N
SC4-31
30 - 100,000
.34N
SC4-34
60 - 200,000
.28N
SC4-81
3 -
10,000
1.29N
SC4-82
3 -
10,000
1.29N
SC4-83
11 -
38,000
1.29N
SSA and Ther-
mosel Spindle
Viscosity (cP)
Shear Rate
sec-1
DV1MRV
DV1MHA
DV1MHB
SC4-14
1,250 -
4,165,000 2,500 -
8,330,000 10,000 - 33,360,000
.40N
SC4-15
500 -
1,660,000 1,000 -
3,320,000 4,000 - 13,280,000
.48N
SC4-25
0 -
167,000
100 -
334,000
400 -
1,336,000
.93N
SC4-27
250 -
830,000
500 -
1,660,000 2,000 -
6,640,000
.34N
SC4-28
500 -
1,660,000 1,000 -
3,320,000 4,000 -
3,280,000
.28N
SC4-29
1,000 -
3,330,000 2,000 -
6,660,000 8,000 -
26,640,000
.25N
Thermosel
SC4-81
37 -
10,000 73.0 -
10,000
292 -
10,000
1.29N
SSA
SC4-82
37 -
10,000 73.0 -
10,000
292 -
10,000
1.29N
SSA
SC4-83
121 -
50,000
243 -
50,000
970 -
50,000
1.29N
UL Adapter
UL Spindle
Viscosity (cP)
Shear Rate
sec-1
DV1MLV
DV1MRV
DV1MHA
DV1MHB
YULA-15 or 15Z
1 - 2,000
7 - 2,000
13 - 2,000
52 - 2,000
1.22N
ULA-DIN-Y
1 - 3,800
11 - 5,000
22 - 5,000
85 - 2000
1.29N