A.2 setup, Figure a-4, Figure a-2 – Brookfield DV1 Viscometer User Manual

Page 32

background image

Brookfield Engineering Laboratories, Inc.

Page 32

Manual No. M14-023

A.2 Setup

1. Be sure that the Viscometer is securely mounted

to the Laboratory Stand, leveled and zeroed

with no cone or cup attached and 0% torque

is displayed.

2. Figure A-2 shows a typical water bath setup.

Connect the sample cup inlet/outlet ports

to the water bath inlet and outlet and set the

bath to the desired test temperature. Allow

sufficient time for the bath to reach the test

temperature. The temperature range of the

Sample Cup (CPE-44Y or CPE-44PY) is 0°-

100°C. Brookfield recommends a maximum

temperature of 80°C to allow for direct hand

contact for adjustment of the micrometer ring.

When using the sample cup at temperatures

near 0°C, be careful to avoid frost buildup on

the top surface of the cup; this could prevent a

proper fit with the micrometer ring. Please refer

to the bath manual for the proper selection of

bath fluid and tubing to ensure safe and proper

operation.

3. The Viscometer has been supplied with a special

cone spindle(s), which contains the Electronic

Gap Setting feature. The “CPE” part number

designation on the cone verifies the Electronic

Gap Setting feature.

4. With the motor off, thread the cone spindle

by using the spindle wrench to secure the

viscometer coupling nut (see Figure A-3);

gently push up on the coupling nut and hold

this securely with the wrench. Thread the cone

spindle by hand. Note: Left Hand Threads.

5. Attach the cup, taking care not to hit the cone

with the cup (Figure A-4). There must be no

fluid in the cup.

6. Option: The sample cup is available with an

optional purge fitting. The user can connect

a dry gas line to this and put a blanket of dry

gas over the sample during measurement, if

desired.

Figure A-3

Spindle

Wrench

(CPE) Cone

These surfaces

must be clean!

Coupling Nut

Micrometer

Adjustment

Ring

Do Not hit the

CONE with the CUP!

Figure A-4

Sample

Cup

(CPE-44Y

or

CPE-44PY)

Bath/Circulator

Bath

Inlet

Bath

Outlet

Cup

Outlet

Cup

Inlet

Micrometer

Ring

Figure A-2