I.5 Installation

Note: “IQ, OQ, PQ”, an abbreviated guideline document for installation, operation and

performance validation for your DV1 digital viscometer can be downloaded from

our website www.brookfieldengineering.com. A more detailed IQ,OQ,PQ procedure

is available for purchase from Brookfield or your authorized dealer.

1) Assemble the Model G Laboratory Stand (refer to assembly instructions in Appendix H).

2) Attach the viscometer head to the clamp on the lab stand rod.

3) Connect the RTD probe to the socket on the rear panel of the DV1.

4) The Viscometer must be leveled. The level is adjusted using the two leveling screws on the

base. Adjust the leveling screws so that the bubble level on the front of the DV1 is centered

within the circle.

Note: Check level periodically during use.

5) Remove the shipping cap which secures the coupling nut on the Viscometer to the pivot cup.

For Cone/Plate Models, hold the Sample Cup and swing the tension bar away from the

bottom of the cup. Lower the cup and remove the foam insert. (Save for future shipments.)

6) Optional: Install the screen protector per the instructions on the package. Additional

installation help can also be found on our YouTube channel: www.youtube.com/user/

BrookfieldEng

Refer to Figure I-2 for the following, except where noted.

7) Make sure that the AC power switch at the rear of the DV1 is in the OFF position.

Connect the power cord to the socket on the back panel of the instrument and plug it into

the appropriate AC line. For Cone/Plate Models, be sure that the toggle switch, used to

activate the electronic gap, is to the left position. (Left when facing the viscometer.)

Note: The DV1 must be earth grounded to ensure against electronic failure!!

Note: The AC input voltage and frequency must be within the appropriate

range as shown on the nameplate of the viscometer (see Section I.2).

8) Turn the power switch to the ON position and allow the viscometer to warm up for 10

minutes before performing autozero.

9) For Cone/Plate models, refer to Appendix A.

10) If appropriate, connect the USB cable (DVP-202) to the USB B port for connection of

the DV1 to the PC.

11) If using the optional Dymo Label Writer 450, connect the cable supplied with the printer

to the USB A port.