to scroll to the required value. Press the SELECT Hot Key to accept the change. Press the BACK

or HOME Hot Key to cancel any change.

II.8.5 Tmp. Offset

The DV1 Viscometer supports the use of an offset for the temperature measurement through the

DVP-94Y temperature probe. The Brookfield DVP-94Y temperature probe has an accuracy of +/-

1.0°C in the range of -100° to 149.9°C and +/-2.0°C in the range of 150° to 300°C. An offset may

be entered into the DV1 to adjust the temperature reading to be in agreement with a third party

measurement device. The offset can be set to any value between -5.0° to 5.0°C or -9.0° to 9.0°F.

TIP: The determination of a temperature offset requires the use of a third party calibrated

temperature probe. Be sure to allow for proper thermal equilibrium between the DVP-

94Y Brookfield probe and the third party probe.

The TMP. OFFSET screen in the Options Menu displays the current offset value. The factory setting

is 0.0°C. To change the TMP. Offset value, press the SELECT Hot Key and then use the Arrow Keys

to scroll to the required number. Press the SELECT Hot Key to accept the change. Press the BACK

or HOME Hot Key to cancel any change.

II.8.6 Language

The DV1 Viscometer supports several languages for the instrument display. Available languages

include: Chinese, English , French, German, Japanese, Portuguese, Spanish, and Russian. The Language

screen in the Options Menu displays the currently selected Language. To select a new language, press

the SELECT Hot Key and then use the Arrow Keys to scroll through the list of available languages.

Press the SELECT Hot Key to accept the language choice. The screen will return to the Options

Menu with the newly selected language indicated.

II.8.7 Connect

The Connect function resets the USB A (printer) and the USB A (computer) communication port.

Use this function to trouble shoot any performance issue when trying to print to the Dymo printer

or communicate with Brookfield Wingather SQ software.
The Connect screen in the Options Menu displays a reminder to check the cable connections as the

first step in trouble shooting. Once the SELECT Hot Key is pressed, the DV1 will take several seconds

to reset the ports and then a message will be displayed indicating that the function is complete. Press

the HOME Hot Key to return to the main screen.

II.8.8 Reset

The Reset function will return all user set parameters (except spindle and speed) to the factory default

settings. The Reset screen in the Options Menu indicates the reset action. Press the SELECT Hot

Key to initiate the reset function. Upon completion of the reset function, the display will be set to

the About Options screen (Section II.8.9).

TIP: The Reset function cannot be reversed.

II.8.9 About

The About screen presents the information seen on the start up screen including Viscometer Name,

Viscometer Model (LV, RV, HA, HB or other) and the firmware version of the viscometer operating

system.