Brookfield Engineering Laboratories, Inc.

Page 21

Manual No. M14-023

TIP: Slow speeds require longer time for equilibrium of the measurement sensor. When using

speed values less than 5 RPM, consider a time value of 60 seconds or longer.

The DV1 Viscometer can be set to infinite time by choosing Time as the end condition and setting

the time value to zero; 00 Hours : 00 Minutes : 00 Seconds. In this configuration, the test will run

continuously until the operator presses the Motor Off button.

Run Until Torq. requires that a % Torque value be selected. % Torque is adjusted by using the Arrow

Keys. The measured viscosity at the defined % Torque (based on the selected spindle and speed)

will also be displayed. The Torq. value is accepted by pressing the SELECT Hot Key. Brookfield

recommends that data be collected between 10 - 100% Torque. Data collected at less than 10% is

not guaranteed to be accurate to the stated instrument accuracy (see Section I.4).

TIP: When measuring low viscosity materials at high speeds of rotation, the beginning of the

test may include large swings in measured torque. Such swings in torque could cause

the Torq. end condition to be met and result in the premature end of the test.

Run Until Temp. requires that a temperature value and tolerance value be selected. Temperature

will be entered in units of C or F as defined by the user (see Section II.8). Temperature and Tolerance

are adjusted by using the Arrow Keys. The SELECT Hot Key is used to advance from Temperature

to Tolerance and finally to accept the input values. The BACK Hot Key can be used to move back

from Tolerance to Temperature.

The Tolerance value provides flexibility in the test method. The temperature measurement system

of the DV1 Viscometer utilizes a Brookfield DVP-94Y temperature probe which has an accuracy of

+/-1.0°C in the range of -100° to 150°C. The accuracy value could result in a measured/displayed

temperature which is not exactly equal to the set Temperature for the Run Until Temp. end condition.

For example, if the temperature probe is placed in a water bath that is set to rise to 50°C, the DV1

Viscometer may display 49.5°C when the temperature bath is at thermal equilibrium. Since the

accuracy of temperature measurement is +/-1°C, this result is considered correct. However, if the Run

Until Temp. end condition is set to 50.0°C, then the test condition will never be met. The tolerance

value allows the instrument to consider the end point achieved when the measured temperature is

within the specified amount of the set point. For example; the Temperature end condition is set to

50.0°C and the tolerance is set to +/-1.0°C. When the measured temperature reaches 49.0°C, the

end condition will then be met and the test will stop.

TIP: Brookfield recommends that the Tolerance value be set to 0.2°C or higher for every Run

Until Temp. end condition.

II.8.2 Test View

The DV1 Viscometer provides the user with two controls for the configuration of the Home Screen.

Examples of TEST VIEW settings are as follows:

1. The user can choose which parameter is displayed on the top line of the display (the primary

parameter). This value will use the largest font size possible. The primary parameter can be set

to any of the following: Viscosity, Time, Temperature, and Torque. The remaining parameters

will be displayed in a smaller font on the second and third lines of the display.

2. The user can choose the mode of how the remaining parameters are viewed, either Static or

Variable.

The Static selection will present a fixed display with all parameters shown simultaneously on

the top three lines of the display. Figure II-7 is an example of a test in progress showing the