Appendix b - warranty repair and service – Brookfield CAP1000 Viscometer User Manual
Page 27
Brookfield Engineering Labs., Inc.
Page 2
Manual No. M02-32A0607
Viscometer Related Variables
• Most fluid viscosities are found to be non-Newtonian. They are dependent on Shear Rate and
the spindle geometry conditions. The specifications of the Viscometer cone and plate geom-
etry will affect the viscosity readings. For example, if one reading is taken at 750 rpm, and a
second at 900 rpm, the two viscosity values produced may be different because the readings
were made at different shear rates. The faster the spindle speed, the higher the shear rate.
• The shear rate of a given measurement is determined by the rotational speed and the cone
angle.
• A repeatable viscosity test should control or specify the following:
1. Viscometer model
2. Cone used
3. Test temperature
4. Test speed [or the shear rate]
5. Length of time to record viscosity
6. Sample volume sufficient to cover the face of the cone
APPENDIX B - Warranty Repair and service