beautypg.com

Brookfield CAP1000 Viscometer User Manual

Page 2

background image

ADDENDUM

Brookfield CAP Viscometer Operating Instructions

CAP1000+ Manual No. M02-312A0607

CAP2000+ Manual No. M02-313-F0211

Addendum No. D12-3312

The tables listed within Sections III.1: Full Scale Range and Accuracy of Measurement and III.2: Ac-

curacy of Viscosity and Temperature of the manuals have been updated, to reflect the recommended

operational speeds for the CAP1000+ and CAP2000+ viscometers. Please refer to the tables within this

addendum.

Table III-1 Full Scale Range Viscosity of LOW TORQUE CAP Viscometer

Cone

Number

Cone Constant

Range

Shear Rate

Constant

FSR at

100 RPM

FSR at

500 RPM

FSR at

any RPM

01

1875

13.33N

0.83

0.17

1875/(22.7*N)

02

3750

13.33N

1.65

0.33

3750/(22.7*N)

03

7500

13.33N

3.30

0.66

7500/(22.7*N)

04

15000

3.33N

6.61

1.32

15000/(22.7*N)

05

30000

3.33N

13.22

2.64

30000/(22.7*N)

06

75000

3.33N

13.04

6.61

75000/(22.7*N)

07

3150

2N

1.39

0.28

3150/(22.7*N)

08

12500

2N

5.51

1.10

12500/(22.7*N)

09

50000

2N

22.03

4.41

50000/(22.7*N)

10

5000

5N

2.20

0.44

5000/(22.7*N)

Table III-2 Full Scale Range Viscosity of HIGH TORQUE CAP Viscometer

Cone

Number

Cone

Constant

Range

Shear

Rate

Constant

FSR

at

100 RPM

FSR

at

400 RPM

FSR

at

750 RPM

FSR

at

900 RPM

FSR

at

any RPM

01

1875

13.33N

18.75

4.69

2.50

2.08

1875/(N)

02

3750

13.33N

37.50

9.38

5.00

4.17

3750/(N)

03

7500

13.33N

75.00

18.75

10.00

8.33

7500/(N)

04

15000

3.33N

150.00

37.50

20.00

16.67

15000/(N)

05

30000

3.33N

300.00

75.00

40.00

33.33

30000/(N)

06

75000

3.33N

750.00

187.50

100.00

83.33

75000/(N)

07

3150

2N

31.50

7.88

N/A

N/A

3150/(N)

08

12500

2N

125.00

31.25

N/A

N/A

12500/(N)

09

50000

2N

500.00

125.00

N/A

N/A

50000/(N)

10

5000

5N

50.00

12.50

6.67

5.56

5000/(N)

(Continued)