Brookfield CAP1000 Viscometer User Manual
Page 13
Brookfield Engineering Labs., Inc.
Page
Manual No. M02-32A0607
Note: 1. CAP Viscometers allow for only one cone at a time of the same cone number to
be calibrated to the viscometer. Multiple cones of the same cone number must each
be calibrated to the viscometer before operation (refer to Appendix B).
2. A special feature of the CAP series viscometers allows the user to perform a cone
calibration with viscosity standard fluids.
3. Cones entered as 11 through 99 must be first calibrated following the directions in
Section III.
If you are not going to calibrate the cone, continue by pressing the
NO key, then the ENtER
key. The viscometer will display the MAIN SCREEN (Figure II-1).
If you are going to calibrate the cone, press the
YEs key, refer to Section III, and follow the
calibration instructions under
Cone Calibration.
II.3 speed setting
The CAP 1000+ is a single speed viscometer. It is supplied from the factory in two configura-
tions: standard and custom. The speed of rotation is shown in the lower right corner of the
display.
Standard
The CAP 1000+ Viscometer is supplied with two speeds: 750 RPM and 900 RPM. All CAP
1000+ Viscometers are factory-set at 900 RPM. To change the setting, press and hold the
stOP
key during the power up sequence. Choose “1=SPEED (RPM)” from the special functions
screen and select the required speed. This selection will become the default speed.
Custom
The CAP 1000+ Viscometer is configured to operate at one speed as specified at the time of
order. This speed is indicated on the instrument serial tag.
To begin spindle rotation, press the
RUN key.
II.4 temperature Control setting
Press the TEMP key and the current temperature setting will blink. The default temperature on
start-up is 25.0°C on low temperature models and 50.0°C on high temperature models.
The temperature ranges are:
Low temperature: 5°C (or 15°C below ambient, whichever is higher) to 75°C
High temperature: 50°C to 235°C
Use the number keys to type in the required set point. The temperature can be set in increments
of 0.1°C. You can turn off the heater by entering 0 on the high temperature unit only.
Use the
ENtER key to accept the new set point.
Note: Thermal equilibrium of the sample and of the spindle must be considered for best
measurement results. Upon powering up the Viscometer or after changing the tem-
perature set point, allow sufficient time for the plate to reach the desired temperature.