Brookfield CAP1000 Viscometer User Manual
Page 15
Brookfield Engineering Labs., Inc.
Page 3
Manual No. M02-32A0607
A maximum of 999 seconds can be printed when running in manual TIMER mode (00). Over
999 seconds will print EEE.
The TIMER (SEC) column will indicate the accumulated time of running at the moment the
print key is pressed while the cone is rotating. This time value will not include the hold time.
At the end of a test, data will automatically be sent to the printer port.
II. Run and stop Keys
The
RUN key has three functions:
1. Press RUN to begin a test.
2. Press and hold the
RUN key for continuous rotation when 00 is the timer setting.
3. Used in executing a cone calibration.
The
stOP key has three functions:
1. Stops the cone rotation at any time.
2. Pressing and holding the
stOP/EsCAPE key during power up selects the special func-
tions mode where the viscosity display units and speed of rotation may be changed.
(Section II.1)
3. Pressing and holding the
stOP and PRINt keys simultaneously executes the printing of
a new heading (Section II.7).