Volume calibration, Start, Service mode – ANKOM TDF Dietary User Manual
Page 84

Operator’s Manual
pg. 84
Rev E 12/2/14
3. Press the
Volume Calibration button. The following screen will be displayed.
4. Label and weigh six small empty glass beakers.
5. Press the
START button on the Volume Calibration screen above. The following screen will be displayed.
←
Place spare containers under IDF Delivery
Nozzles to catch pre-calibration charge.
Press OK when done.
OK
Volume Calibration
Volume Calibration
←
xx.x
Calibration Time
START
←
Service Mode
Temperature
Sensor
Corrections
IDF Count
Volume
Calibration
Bypass Amylase
Phase Initial
Heat-up
(NO)
AOAC 2009.01/
2011.25
Registration