Motor Test
When you press the
Motor Test button on the Diagnostics screen, the following screen will be displayed.
This diagnostic allows you to test the motor at different speeds (default is 240 RPM) and directions using different amounts of
fluid. Before pressing the
GO button to start the Motor Test, it is recommended that you first open at least one Supply Valve and
one Output Valve. To do that, press the
Set Valves button. The following screen will be displayed.
This diagnostic also displays the maximum pressure read from the vacuum sensor during the test. The value starts at zero and
will display a value while the motor is running and when it stops. This can be used for troubleshooting.
Running the Motor Test could leave fluid in some of the lines. After running a
Motor Test, always run the Clear IDF/SDF Lines diagnostic.
←
Set Valves
Amylase Supply
(closed)
HCl Supply
(closed)
AMG Supply
(closed)
Protease Supply
(closed)
Water Supply
(closed)
EtOH95 Supply
(closed)
EtOH78 Supply
(closed)
Buffer Supply
(closed)
IDF Inlet
(open)
Waste
(open)
SDF Inlet
(open)
Output
Supply
←
Motor Test
Set Valves
GO
Motor
Direction
(FWD)
120 RPM
240 RPM
Motor Speed
10.0
Amount (ml)
00.0
Max Vacuum (psi)