Calibrate the module

1. Click within the module row, and then click Calibrate to activate the test sequence for the module.
2. Enter the information about any replacement parts you used during service of the product. After entering the part

information, click Begin to start the tests.

3. Follow the onscreen prompts in the right side pane of the test window to progress through the test sequence. The

left detail pane shows the order of the required tests and tracks the test sequence progress.
As the Welch Allyn Service Tool application completes each test, the pass/fail indicator changes to green for pass

or red for fail, and the application proceeds to the next test.
To terminate the active test sequence at any time, clickAbort. To terminate all tests, clickAbort All.
After the Welch Allyn Service Tool application completes all the tests, it displays the pass/fail test results. After

you have finished reviewing the test results, clickClose.