Species

Code Definition /

Description

Original / Alert

type

Traditional code

map

Details

Host

controller

action

6

19

Host controller

alert

Internal error

None

None

6

20

Host controller

alert

Internal error

None

None

6

21

Host controller

alert

Internal error

None

None

6

24

Host controller

alert

Internal error

None

Request

appropriate

CIT.

NIBP reading status codes

Code Definition /

Description

Origin / Alert

type

Traditional code

map

Display message to

operator

Host controller

action

0

No error.

Not applicable. Not applicable.

Not applicable.

Display the

reading.

1

Reading aborted.

Operator alert. C01

Blood pressure reading

cancelled by user.

Return cancelled

cycle confirmation

to operator.

2

NIBP reading not

available.

Host controller

alert.

Internal error.

None

Host controller to

resign attempt to

gather reading.

3

Reading may be

inaccurate.

Patient alert.

Not applicable.

Message displayed

depends on host

environment. Message

could be result of a

reading error, or could be

the reading accompanied

by a splat (*).

Make the error

clearable, and

allow another

cycle to be run.

4

Excessive artifact,

no BP parameters

could be calculated.

Patient alert.

C05 / C06

Unable to determine

blood pressure. Check

connections; restrict

patient movement.

Make the error

clearable, and

allow another

cycle to be run.

5

Low inflation.

Patient alert.

C05

Unable to determine

blood pressure. Check

connections; restrict

patient movement.

Make the error

clearable, and

allow another

cycle to be run.

6

Kinked tube

detected or there is a

transducer

calibration failure.

Equipment

alert.

Internal error.

Internal NIBP error.

Contact Welch Allyn

Technical Service.

Power down.

7

Dump too soon.

Patient alert.

C05

Unable to determine

blood pressure. Check

connections; restrict

patient movement.

Make the error

clearable, and

allow another

cycle to be run.