Welch Allyn Connex Vital Signs Monitor Calibration/Verification Module Notes - Quick Reference Guide User Manual
Page 5
Species
Code Definition /
Description
Original / Alert
type
Traditional code
map
Details
Host
controller
action
6
19
Host controller
alert
Internal error
None
None
6
20
Host controller
alert
Internal error
None
None
6
21
Host controller
alert
Internal error
None
None
6
24
Host controller
alert
Internal error
None
Request
appropriate
CIT.
NIBP reading status codes
Code Definition /
Description
Origin / Alert
type
Traditional code
map
Display message to
operator
Host controller
action
0
No error.
Not applicable. Not applicable.
Not applicable.
Display the
reading.
1
Reading aborted.
Operator alert. C01
Blood pressure reading
cancelled by user.
Return cancelled
cycle confirmation
to operator.
2
NIBP reading not
available.
Host controller
alert.
Internal error.
None
Host controller to
resign attempt to
gather reading.
3
Reading may be
inaccurate.
Patient alert.
Not applicable.
Message displayed
depends on host
environment. Message
could be result of a
reading error, or could be
the reading accompanied
by a splat (*).
Make the error
clearable, and
allow another
cycle to be run.
4
Excessive artifact,
no BP parameters
could be calculated.
Patient alert.
C05 / C06
Unable to determine
blood pressure. Check
connections; restrict
patient movement.
Make the error
clearable, and
allow another
cycle to be run.
5
Low inflation.
Patient alert.
C05
Unable to determine
blood pressure. Check
connections; restrict
patient movement.
Make the error
clearable, and
allow another
cycle to be run.
6
Kinked tube
detected or there is a
transducer
calibration failure.
Equipment
alert.
Internal error.
Internal NIBP error.
Contact Welch Allyn
Technical Service.
Power down.
7
Dump too soon.
Patient alert.
C05
Unable to determine
blood pressure. Check
connections; restrict
patient movement.
Make the error
clearable, and
allow another
cycle to be run.